
4
Place each BeadChip into a black frame, aligning the barcode with the alignment fixture ridges. Fully
submerge each BeadChip.
5
Place a
clear
spacer onto the top of each BeadChip. Use the alignment fixture grooves to guide the
spacers into position.
The white spacers are not a substitute for the clear spacers.
6
Place the alignment bar onto the alignment fixture. Fit the groove on the alignment bar over the tab on the
alignment fixture.
7
Place a clean glass back plate on top of each clear spacer. Position the plate reservoir at the barcode
end of the BeadChip, facing inward to create a reservoir against the BeadChip surface.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
23
Infinium HD Super Assay Reference Guide