66
Ending CC Temperature control
The temperature control can be ended at any time by pressing the function key T3 (65).
The temperature control and circulation is immediately stopped. Alternatively, to set the
unit to standby mode and use the main menu point
Start & Stop
- menu point
Stop
temperature control.
For units with compressor:
The main switch (36) can be used to turn the power off only after the compressor has
been stopped by the controller.
Room temperature should be reached before the temperature control is ended.
Protect your application by
NOT
closing any isolation and drain valves. Overpressure /
underpressure can be built up within the system if valves are closed and temperature
adjustment takes place.
Filling an externally closed system
Caution!
-
Fill to the unit to the minimum level necessary.
-
Please refer to local regulations and internal procedures.
-
When filling the unit, extra precautions such as earthing the expansion vessel,
fluid container funnel and application may be necessary.
-
Personal Protection Equipment (PPE) should be worn as required by the fluid
MSDS sheets, and local regulation.
-
Please note the temperature of the thermal fluid. The fluid should be left at room
temperature for a few minutes before draining.
Warning!
Overflowing thermal fluid will create a film on surfaces, which should be cleaned up and
properly disposed of as soon as possible in accordance with the MSDS information. If
thermal fluid is spilled over the unit, the unit should be immediately turned off, and
Huber-trained personnel consulted.
Failure to observe the above precautions may mean that the unit will not comply with
all of the requirements of DIN EN 61010-2-010.