21
Field 10: Display delta T
This field displays the delta T value (max. admissible difference between process and
internal temperature). This value may be set within a range of 0…100K under the main
menu point limits / delta T limits. This field is active only with a connected process
sensor and when the temperature control mode
process temperature
is activated.
Field 11: Display Over-temperature cut-off
This field displays the current setting of the over-temperature cut-off. Please note that
this value can only be changed through the
Main menu
Over-temperature.
Please refer
to the
Setting the over-temperature
chapter in the
Main menu
.
Field 12: Display Alarm and Warning messages
This field displays information on any alarm or warning conditions that are present.
Alarm and warning messages are also immediately displayed as text in the graphic
display (60).
Field 13: Display Date and Time
This field displays the current date and time.