31
Example: Indication when choosing
Large display
Enter program
This corresponds to Function F20 in the Funct.-no. menu.
Here it is possible to write new programs, or programs already written can be edited
and changed or erased. (
add segments, insert segments, delete segments or edit
segments
) or erase whole programs.
Also one can set a particular behaviour at the end of the program through
Stop temp.
control, Continue temperature control
(temperature is continued at the last set-point) or
Repeat
(the temperature program is restarted). One can also display the program
elements as text or graphic. Working with the program creator will be described below.
L
L
e
v
e
l
TInternal °C
-20.5
TProcess °C
-20.1
Tset-point °C
-20.0
OT 35 °C
Temperature control is
active