37
Enter program
A description on this menu point can be found in chapter
Compact menu.
Factory default
This section allows the different areas of the temperature control unit to be reset to the
factory default. This can be a relatively quick way of changing the unit settings.
Unit control data:
Resets the set-points, set-point limits, temperature control mode, to the factory-set
default values. Settings in the user menu and programs created using the programmer
remain unchanged.
User menus:
Resets the complete user menus to their default settings. Settings in the unit data and
programs created using the programmer remain unchanged.
Programmer:
Resets complete programs to default settings. Settings in the unit data and user menus
remain unchanged.
All together:
Resets the unit data, user menu, program, and controller parameters to default values.
Language
This corresponds to Function F90 in the Funct. No. menu, and allows the unit’s
operating language to be selected. The language options displayed are available.
Limits
The following functions are available:
1.
Delta T limit (limitation of the jacket temperature to the reactor temperature)
2.
Maximum heating power (limitation of the heating power in % steps)
3.
Maximum cooling power (limitation of the cooling power in % steps)
4.
Go back
You can here set the maximum allowable difference (
Delta T limits)
between the
internal temperature (jacket temperature) and the process temperature when using
process control. If the chosen temperature difference is reached, then the temperature
control device power is reduced so that this temperature difference is held. This
function can protect the application (e.g. glass reactor) against thermal stress caused
by too high a Delta T.
Over-temperature protection (valid for units with heating)
A description on this menu point can be found in the chapter on
Compact menu.
Program start & stop
A description on this menu point can be found in chapter
Compact menu.