35
Comfort menu
Acoustic alarm
Here you have the option to activate / deactivate the acoustic signal output.
Auto-Start (after power on)
This corresponds to Function F5 in the Funct.-no. menu. This allows the start-up
condition, after mains failure to be defined.
The following is true:
OFF
/ Standby
After power loss
Temperature control will
not
be restarted when power restored
(Default setting)
ON
/ Temp. control active
After power loss
Temperature control will be restarted on return of power
Power failure automatic
After power loss
Return on power the settings previously set will be taken over
Caution!
The end-user should assess the risk and consequences for their application and select
one of these three functions. The default setting is
OFF
.
Clock
Sets the unit Time and date
.
A number of functions can be chosen, e.g. a calendar /
reminder function and timed start can also be configured.
Example:
Set alarm clock
First enter the temperature programme via the main menu point
clock / alarm clock
function / acoustic signal
. The acoustic signal will be given out when the time (date) is
set via the function
clock
/
action alarm clock
/
set alarm clock.
CLOCK FUNCTIONS:
Set time
Set date
Autom. summer time
Set alarm clock
Alarm clock function
Return to main menu
Date :
13.05.08
Time:
09:39:30
Alarm clock date: 18.05.00
Alarm clock time: 04:30:00
Action:
Program 1 Start
FUNCTION ALARM CLOCK:
No function
Acoustic signal
Program start
Set alarm clock
Return to main menu