88
Exchange of the CC Electronics / Remote Control
Disconnect the thermostat from mains power by turning the main switch (36) to
OFF.
Pull out the power plug. In the case of a problem with the electronics, you can
exchange these parts yourself or prepare it for use as a remote control. For questions or
difficulties please contact our representatives or ourselves.
-
Remove the lock (anti-clockwise) for securing the CC-Pilot on the front side of
the CC-Pilot, and pull the Pilot carefully up and away.
-
Place the replacement CC-Pilot carefully in place and fasten it with the locking
(clockwise)
Remote Control:
An unrigged CC-Pilot Controller can also be used as a remote control. Therefore, mount
the connection wire (order no. #16160) between the machine and the CC-Pilot.
-2048
Internal actual value
has
exceeded alarm temperature
Check alarm limits
-2049
-2050
-2051
-2052
-2053
-2064
-2068
-2072
-2075
-2095
Internal actual value
has fallen
below alarm temperature
Process actual value
has
exceeded alarm temperature
Process actual value
has fallen
below alarm temperature
Analogue input
is selected.
Signal is faulty. Under/over scale
Analogue input
is selected.
Signal is faulty. Near limit
Pump pressure
is too high
Temperature difference
control
sensor, and OT sensor too high
Over-temperature
in expansion
tank almost reached
Actual value
lies outside
set-point range
Caution:
Level too low. Switch
off will take place shortly.
Check alarm limits
Check alarm limits
Check alarm limits
Please note chapter on: analogue interface
Please note chapter on: analogue interface
Flow resistor is too high. Improve thermo regulation by
e.g. better cross-section enlargement of the hoses
Improve thermo regulation by e.g. higher pump speed,
enlarge the cross-section of the hoses
Please note chapter on „setting the over-temperature”
Check set-point limits
Possibly insufficient thermal fluid in the circuit. Fill in
thermal fluid.