50
F8 POCO - programming
The options for the potential free contact are given and described in the earlier
Potential
free Contact section the Com.G@te menu
of this manual
(Pot. free Contact).
F9 Control parameters
Please see chapter
Control parameters
in the
Compact menu
for detailed description.
F10 Machine messages
Information on the machine about condition (status, warnings and faults).
F12 Adj. internal sensor (Adjust internal sensor)
Up to 5 free selectable temperature values for the adjustment of the internal sensor can
be defined and adjustment may be carried out. Please also see description on
Sensor
adjustment
in the chapter
Comfort menu.
F13 Adj. process. sensor (Adjust process sensor)
Up to 5 free selectable temperature values for the adjustment of the process sensor can
be defined and adjustment may be carried out. Please also see description on
Sensor
adjustment
in the chapter
Comfort menu.
F14 Adj. return sensor (Adjust return sensor)
Up to 5 free selectable temperature values for the adjustment of the return sensor can
be defined and adjustment may be carried out. Please also see description on
Sensor
adjustment
in the chapter
Comfort menu.
F18 Delta T limit
Maximal admissible temperature difference between internal and process temperature.
Once the maximum temperature difference has been reached, the unit will automatically
reduce its cooling (or heating) capacity as required.
F19 Ramp function
The set-point default refers to, depending on the temperature control mode set
(function F3) the internal sensor or process sensor.
F20 Enter program
You can enter the chosen temperature programme.
F22 Program control
Choose between following options:
Start, Stop, Break, Skip
to the end segment
of a
running temperature programme.
F23 Program start
Start of the temperature programme (calendar start).
F27 Time scale
Time scale in minutes or hours.
F28 Ext. control signal (External control signal)
The external control signal can be used to control one of a number of available unit
functions. Please see chapter on
Com.G@te menu.