76
Specific functions Com.G@te
(internal version)
1
2
3
6
5
4
3
2
1
3
2
1
1
5
9
6
(external version)
3
1
2
1
3
2
1
6
5
4
3
2
3
1
2
1
5
9
6
The signal connectors have been designed according to the NAMUR standard
AIF Reg-E-Prog Socket
The analogue interface, one input channel (programmable, option current or voltage
signal possible) and 3 output channels (1 programmable).
The analogue interface of the Com.G@te is programmed in the
Main menu
. Please also
read the description concerning the analogue interface in the section of the
Com.G@te menu.
Pin
Signal
1. Current output, T external
0/4-20mA oder 0-10V
2. Current output, set-point 0/4-20mA
oder
0-10V
3. GND for analogue outputs
GND
4. Analogue input (programmable)
0/4-20mA oder 0-10V
5. Current output, free programmable
0/4-20mA oder 0-10V
6. GND for analogue input
GND
Further to the connections described under Com.G@te (internal version) the Com.G@te
(external version) offers an external level monitoring (LEVEL) connection.
Level Socket
For level monitoring in the sight glass.
This connection offers the option of connecting an external float switch (Order No.
#6152), which is fitted in the sight glass in order to monitor the level of your external
closed application.