85
Alarm and Warning codes
Hard Alarms (not resettable)
Code
Message
Action
-1
Please see chapter on setting the over-temperature
Over
-temperature reached
-2
Low-level alarm release
Fill with sufficient thermal fluid
-4
-5
-6
-7
-9
-10
actual value internal
has
exceeded the alarm temperature
actual value-internal
has
exceeded the alarm temperature
actual value process
has
exceeded the alarm temperature
Process actual value
has fallen
below alarm temperature
PT100 Internal
not connected or
faulty
PT100 Process
not connected or
faulty
Check alarm limits
Check alarm limits
Check alarm limits
Check alarm limits
Please contact us or our service partners.
Connect a properly functioning external PT100.
-11
-14
-15
-42
-43
-48
-49
-50
-58
PT100 Condensation
not
connected or faulty
Condensation
pressure or
temperature too high
Pump speed
not correct. Is
pump running?
PT100 return sensor
not
connected or faulty
PT100 Evaporator end
not
connected or faulty
Evaporation pressure
has fallen
below allowed minimum value.
Evaporation pressure
is too high
despite controlling
Superheat
is too low for a period
Standby current
is too high
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
Please contact us or our service partners.
-59
Current
too high when pump is
running
Please contact us or our service partners