13
Environmental Conditions
This unit, and operations, will comply with DIN EN 61010-1:2001, only when it is
located in suitable environmental conditions.
-
for indoor use only;
-
installation site
≤
2000 m altitude;
-
installed on a level, even, non flammable surface;
-
maintain a clearance above and around the unit of 10 cm for water-cooled units,
and 20cm for air-cooled units, to allow air to circulate around the unit;
-
for ambient temperature conditions please refer to the technical data sheet;
remaining within these ambient conditions is imperative in ensuring accurate
operation;
-
maximum relative humidity of 80% up to 32°C, decreasing linearly to 50%
relative humidity at 40°C
-
use only as long a power cord as necessary;
-
the unit should be located so as not to restrict access to the mains power
switch;
-
mains voltage should be ±10% of the rated value;
-
avoid voltage spikes;
-
transient voltage surges as they occur normally in the supply grid;
-
clean rating 2;
-
overvoltage category II