TROUBLE SHOOTING
286
Service Manual
Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Warnings
Numbers in the range 6272-6335 are warning codes. Warnings will set
the status of the Primary value to UNCERTAIN.
Status code
(SmartView)
Description
(Engauge)
Remarks
6273
No valid product level for
compensation
E.g. because the displacer is not
measuring product level. Please check
the Primary value of the Product level.
6274
No valid temperature for tank shell
compensation
One of the temperatures needed for
the tank shell compensation is not
available (refer to section
7.7.2.3
).
Please check the vapour and product
temperatures on the FCI-HRT or
FII-RTD.
6275
Balance test number of valid test
points too low
The unbalance table has too few valid
entries. Can appear when the
unbalance compensation for set point
or density is enabled. Perform a
balance test to provide a new table
(refer to sections
7.7.1
and
9.3.1
).
6276
Balance test number of valid test
points too high
The unbalance table has too many
valid entries. Can appear when the
unbalance compensation for set point
or density is enabled. Perform a
balance test to provide a new table.
(refer to sections
7.7.1
and
9.3.1
)
6277
No valid vapour temp. for wire
temp. compensation
Please check the vapour temperature
on the FCI-HRT.
6278
No valid magnet temp. for magnet
temp. compensation
Please check the magnet temperature
status (Board Specific).
6279
No valid product level for
hydrostatic deformation comp.
E.g. because the displacer is not
measuring product level. Please check
the Primary value of the Product level.
6280
No valid displacer temp. for
volume compensation
Please check the product temperature
on the FCI-HRT.
6283
No valid product level or magnet
temp. for drum temp.
compensation
Please check the magnet temperature
status (Board Specific) or the Primary
value of the Product level.
6285
RIC configuration parameters not
commissioned
6286
Product level below RIC minimum
level
6287
Error getting HART dynamic
variables entity device address 11
RIC warning
6288
Error getting HART dynamic
variables entity device address 12
RIC warning
6289
Error getting HART dynamic
variables entity device address 13
RIC warning
6290
Wrong HART units device address
11
RIC warning
6291
Wrong HART units device address
12
RIC warning
Содержание FlexLine SmartServo 954
Страница 1: ...Service Manual 954 SmartServo FlexLine ...
Страница 197: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 189 954 SmartServo FlexLine ...
Страница 202: ...COMMISSIONING 194 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Страница 203: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 195 954 SmartServo FlexLine ...
Страница 300: ...APPENDIX A 292 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Страница 302: ...APPENDIX B 294 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...