TROUBLE SHOOTING
Part No.: 4417340 Revision 1
Service Manual
289
954 SmartServo FlexLine
Status code
(SmartView)
Description
(Engauge)
Remarks
6191
Encoder LED pin malfunction
Hardware error. Malfunction of the LED
of the reference encoder.
6208
Wire tension too
high
The wire tension is too high. The gauge
will resume functioning when the wire
tension issue is resolved (by e.g. moving
the displacer down). Refer to sections
7.7.3
and
11.1
.
6209
Wire tension too
low
The wire tension is too low. The gauge
will resume functioning when the wire
tension issue is resolved (by e.g. moving
the displacer up). Refer to sections
7.7.3
and
11.1
6210
Density displace declined
6211
Boot self test failed
6212
Density license not available