COMMISSIONING
Part No.: 4417340 Revision 1
Service Manual
145
954 SmartServo FlexLine
Name
Explanation
Value Range
Default
[Element skipping]
Enables or disables the fact that
elements can be skipped if an
element is outside a temperature
range.
<enable>
<disable>
<disable>
[Maximum skipped elements]
The number of elements that
can be skipped
<1,2>
<1>
[Median filter]
Enables or disables a
median filter of 5 levels deep
(spike filtering).
<enable>
<disable>
<enable>
[Temperature calculation method]
The way the average
temperature is calculated (see
also features above)
<Standard>
<Custom>
<Enhanced>
<Enhanced>
[MRT length table]
[Only for MRT]
Specifies whether a fixed range
of MRT resistors is used (=
<F>
)
or user configured lenghts (=
<T>
). Fixed lengths are:
0.65, 1.25, 1.95, 2.85, 4.15, 5.65
1 character
<F>
<T>
<F>
[MRT element length
used if MRT length table is 'T']
[Only for MRT]
The lengths of the MRTs
including anchor eye
floating point numbers:
<-x.x .. +x.x>
<0.0>
Engauge * Vapour temperature * Tab
NOTE: Some settings for Vapour temperature calculations are
shared via the product temperature settings.
☛
The following entities
can
be set by Engauge or HART
SmartView for a specific functioning of the FII-RTD module in an
instrument. The default values are available after initialization of the non
volatile memory; so check each entity for its correctness.
Содержание FlexLine SmartServo 954
Страница 1: ...Service Manual 954 SmartServo FlexLine ...
Страница 197: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 189 954 SmartServo FlexLine ...
Страница 202: ...COMMISSIONING 194 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Страница 203: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 195 954 SmartServo FlexLine ...
Страница 300: ...APPENDIX A 292 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Страница 302: ...APPENDIX B 294 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...