2-25
3 The measurement is completed in four minutes, and coefficient (calculated value)
will be displayed.
If “Fault” is displayed for “Proc. Status”, follow the instructions given in
“4.12 Troubleshooting when Fault Occurs in Viscometer.”
4 Press OK.
If the OK key is pressed when
“
Proc. Status” is “Ended”, the coefficient will be
registered, and the circulation control screen will be displayed.
2.6.2 Operation test of viscometer
●This function allows you to check the viscometer operation and measure the current ink
viscosity. Note that this operation test is not to determine the coefficient.
● It is not necessary to use new ink for operation test.
Coefficient
OK
[Note]
The coefficient value will not be registered when Cancel is pressed instead of OK key.
[Standard value of the viscometer calibration] 1200 to 2200
*
The value stated above is a reference. There will be no problem even if the coefficient
value displayed is other than stated above as long as data of Plunger falling time keeps
within a similar range.
OK
Cancel
<the previous calibration>
Exec.time 2018/07/07 11:30
< the current calibration >
Exec.time 2018/07/07 12:45
Proc. Status: Ended
Viscometer calibration
HOME
[Stop ]
2018.07.07 12:45
Com=0
Plunger falling time
Plunger falling time
1st
2nd
5th
Average
3rd
4th
Inside temperature XX
( )
Standard X. XXX (mPa
▪
s)
X.XX (sec)
X.XX (sec)
X.XX (sec)
X.XX (sec)
X.XX (sec)
X.XX (sec)
1st
2nd
5th
Average
3rd
4th
Inside temperature XX
( )
Standard X. XXX (mPa
▪
s)
X.XX (sec)
X.XX (sec)
X.XX (sec)
X.XX (sec)
X.XX (sec)
X.XX (sec)
Coefficient XXXX
Coefficient XXXX
Ink viscosity XXX
When updating coefficient, press [OK].
Содержание UX Twin Nozzle
Страница 1: ...Service Manual HITACHI Printer UX Twin Nozzle Model Revision February 2017 Version A ...
Страница 106: ...3 30 Circuit diagram of EZJ127A board ...
Страница 123: ...3 47 Circuit diagram of EZJ129 board ...
Страница 131: ...3 55 3 3 20 Electrical Connection Diagram ...
Страница 151: ...4 17 7 Reset the time of the replaced solenoid to 0 on the Parts usage time management screen ...
Страница 352: ...7 1 7 Attached Drawing 7 1 Circulation System Diagram Circulation System Diagram ...
Страница 354: ...7 3 7 3 Electrical Connection Diagram Electrical Connection Diagram ...
Страница 355: ...7 4 7 4 Dimensions around charge electrode and deflection electrode Nozzle diameter 65μm Nozzle 2 side Nozzle 1 side ...