PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
15/107
3. Command Reference (Device-Specific Commands)
Set and Query Frequency Full Scale
Syntax
Command
:AOUT:FREQuency (Frequency Data)
Query
:AOUT:FREQuency?
Response
Frequency Data :100Hz/500Hz/1kHz/5kHz
Description
Command
Sets
the maximum frequency of D/A output’s Frequency Full Scale and the
motor.
Query
Sets
the maximum frequency of D/A output’s Frequency Full Scale and
returns it as words.
Example
Command
:AOUT:FREQ 100Hz
Set the Frequency Full Scale of D/A output as 100Hz.
Query
:AOUT:FREQ?
Response
:AOUT:FREQUENCY 100Hz (when HEADER is ON)
100Hz (when HEADER is OFF)
Note
The settings for the D/A output’s Frequency Full Scale and the motor measured maximum
frequency are the same.
Set and Query Coefficient of Integrated Full Scale
Syntax
Command
:AOUT:INTEGrate <Magnification data>
Query
:AOUT:INTEGrate?
Response
Magnification Data
:1/10,1/2,1,5,10,50,100,500,1000,5000,10000
Description
Command
Sets the integration full scale coefficient of D/A Output.
Query
Sets the integration full scale coefficient of D/A output and returns it as
words.
Example
Command
:AOUT:INTEG 1
Set the coefficient of the D/A integration full scale as 1.
Query
:AOUT:INTEG?
Response
:AOUT:INTEGRATE 1 (when HEADER is ON)
1 (when HEADER is OFF)