![Hioki PW3390 Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/hioki/pw3390/pw3390_instruction-manual_2137353020.webp)
PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
20/107
Set and Query Efficiency, Pout of Loss Calculation Formula
Syntax
Command
:CALCulate[number]:POUT <P1/P2/P3/P4/P12/P34/P123/Pm>
Query
:CALCulate[number]:POUT?
[number]
・・・・
1,2,3
Response
“P1/P2/P3/P4/P12/P34/P123/Pm”
Description
Command Sets the items for Efficiency, Pout of Loss Calculation Formula.
Query
Returns the setting items for Pout of Loss Calculation Formula as words.
Example
Command
:CALC1:POUT Pm
Set the Pout item of Calculation Formula 1 as Pm.
Query
:CALC1:POUT?
Response
:CALCULATE1:POUT Pm (when HEADER is ON)
Pm (when HEADER is OFF)
Note
When P12/P34/P123 cannot be selected because of wiring settings, they cannot be specified.
Pm cannot be specified except when it can be selected during the implementation of the motor
analysis function.
Set and Query Automatic Saving
Syntax
Command
:CARD:AUTO:SAVE <ON/OFF>
Query
:CARD:AUTO:SAVE?
Response <ON/OFF>
ON: Automatic Save on
OFF: Automatic Save off
Description
Command
Sets the automatic saving to the CF Card On or OFF
Query
Returns the setting for the automatic saving to the CF Card as On or OFF.
Example
Command
:CARD:AUTO:SAVE ON
Set the automatic saving to the CF Card ON.
Query
:CARD:AUTO:SAVE?
Response
:CARD:AUTO:SAVE ON (when HEADER is ON)
ON (when HEADER is OFF)
Содержание PW3390