PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
107/107
5. Troubleshooting
Please refer to the following contents for any operation that is not processed properly.
Symptoms
Causes
Solutions
Cannot
communicate at all
Is the LAN cable connected
correctly?
When using a HUB to connect the main instrument
and the computer, use a straight cable.
When connecting the main instrument and the
computer as 1:1, either attach a cross conversion
connector to the straight cable or use a cross cable
to connect.
Is the Interface setting
correct?
Confirm the settings for the IP address, subnet
mask and default gateway of the main instrument.
Is the IPAddress the same
as other instruments?
Confirm the IPAddress setting of the main and
other instruments.
Is the TCP/IP port number
correct?
Specify the correct port number.
Cannot
communicate well
Is the message terminator
(delimiter) correct?
Insert CR+LF to divide each message that is
forwarded.
Command
not
executed
Is there any error?
Use *ESR? and look at the contents of the
standard event status register to confirm the error.
Can
the
command
be
operated?
Many commands for settings cannot be executed
during hold or integration operation.
No response at all
to the queries sent
Is the response acquired
correctly?
Read after sending every query.
To read all at once, use a message separator and
describe the Query in one line.