Set and Query Interval Time
Syntax
Command
:INTERval <Time data>
Query
:ITERval?
Response
OFF/50ms/100ms/200ms/500ms/1s/5s/10s/15s
/30s/1min/5min/10min/15min/30min/60min
Description
Command
Sets the interval time.
Query
Returns the interval time setting as a character string.
Example
Command
:INTER 1min
Set the interval time to 1 minute.
Query
:INTER?
Response
:INTERVAL 1min (WHEN HEADER IS ON)
1min (WHEN HEADER IS OFF)
Set and Query IP Address
Syntax
Command
:IP:ADDRess <Address 1>,<Address 2>,<Address 3>,<Address 4>
Query
:IP:ADDRess?
Response
Address 1
:000 to 255
Address 2
:000 to 255
Address 3
:000 to 255
Address 4
:000 to 255
Description
Command
Sets IPAddress.
Query
Returns IPAddress setting as a numerical value.
Example
Command
:IP:ADDR 192,168,1,1
Set IPAddress to 192.168.1.1.
Query
:IP:ADDR?
Response
:IP:ADDRESS 192.168.1.1 (WHEN HEADER IS ON)
192.168.1.1 (WHEN HEADER IS OFF)
Note
The address will be changed after the command has been sent. Please be careful when
sending from LAN.