![Hioki PW3390 Скачать руководство пользователя страница 99](http://html.mh-extra.com/html/hioki/pw3390/pw3390_instruction-manual_2137353099.webp)
PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
99/107
Select and Query Voltage Rectifier Type
Syntax
Command
:VOLTage [CH]:MEAN <ON/OFF>
Query
:VOLTage [CH]:MEAN?
[CH]
・・・・
1,2,3,4
Response
ON
: Sets the voltage rectifier type to MEAN.
OFF
: Sets the voltage rectifier type to RMS.
Description
Command
Selects the RMS/MEAN of the voltage rectifier type.
Query
Returns the selection of the RMS/MEAN of the voltage rectifier type as ON
(MEAN) or OFF (RMS).
Example
Command
:VOLT1:MEAN OFF
Select the rectifier type of voltage Channel 1 as RMS.
Query
:VOLT1:MEAN?
Response
:VOLTAGE1:MEAN OFF (WHEN HEADER IS ON)
OFF (WHEN HEADER IS OFF)
Note
By combining measurement lines (for above IP3W), the settings for the voltage rectifier types of
other channels which are combined are also changed.
Содержание PW3390
Страница 108: ......