PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
56/107
Select and Query Motor Analysis Function Channel B Units
Syntax
Command
:EXTernalinB:UNIT <Unit data>
Query
:EXTernalinB:UNIT?
Response
<Unit data>
: V/Hz/rpm
Description
Command
Sets the unit for Channel B.
Query
Returns the unit setting for Channel B as a character string.
Example
Command
:EXTB:UNIT V
Set the Channel B unit to V.
Query
:EXTB:UNIT?
Response
:EXTERNALINB:UNIT V (WHEN HEADER IS ON)
V (WHEN HEADER IS OFF)
Note
During the pulse setting, when Channel B sets to “V”, Channel B input setting changes to
Analog DC.
Select and Query Noise Analysis Measurement Channel
Syntax
Command
:FFT:ITEM <CH1/CH2/CH3/CH4>
Query
:FFT:ITEM?
Response
<CH1/CH2/CH3/CH4>
: Measurement channel character string
Description
Command
Sets noise analysis measurement channel
Query
Returns setting for noise analysis measurement channel as a character
string.
Example
Command
FFT:ITEM CH1
Set noise analysis measurement channel to Channel 1.
Query
FFT:ITEM?
Response
FFT:ITEM CH1 (WHEN HEADER IS ON)
CH1 (WHEN HEADER IS OFF)
Содержание PW3390
Страница 108: ......