PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
76/107
Set and Query Output Order of Harmonic Data Communication Output
Syntax
Command
:MEASure:ITEM:HARMonic:ORDer <Lower limit order>,
<Upper limit order>, <ODD/EVEN/ALL>
Query
:MEASure:ITEM:HARMonic:ORDer?
Response
Lower limit order (NR1)
:0 to 100
Upper limit order (NR1)
:0 to100
ODD
: Odd-number order only
EVEN
: Even-number order only
ALL
: All orders
Description
Command
Sets the default send item of the response data in the default mode as a
response to ":MEASure:HARMonic?" as 0
– 255.
Use in a combination with ":MEASure:ITEM:HARMonic:LIST"Command.
Query
Returns the setting for the default send item of the response data in
the default mode as a response to ":MEASure:HARMonic?" as numerical
values or character strings.
Example
Command
:MEAS:ITEM:HARM:ORD 1,15,ODD
Set the odd-number order from 1
– 15 to the default output.
Query
:MEAS:ITEM:HARM:ORD?
Response
:MEASURE:HARMONIC:ORDER 1,15,ODD (WHEN HEADER IS ON)
1,15,ODD (WHEN HEADER IS OFF)