Command
:STIMe:STOPtime <Year>,<Month>,<Date>,<Hour>,<Minute>
Query
:STIMe:STOPtime?
Response
Year data:2000 to 2079 (can be set 00 - 79)
Month data:01 to 12
Date data:01 to 31
Hour data:00 to 23
Minute data:00 to 59
Description
Command
Sets the actual time stop time.
Query
Returns the time setting of the actual stop time as numerical values of year,
month, date, hour and minute.
Example
Command
:STIM:STOP 08,12,25,12,00
Set the actual time stop time as 2017 Dec 25
th
12:0
Query
:STIM:STOP?
Response
:STIME:STOPTIME 2017,12,30,8,30 (WHEN HEADER IS ON)
2017,12,30,8,30 (WHEN HEADER IS OFF)
Note
The instrument can interpret days of the month as well as leap years, so specifying an
improbable date will lead to an error.
When a time before the start time is set, an execution error will occur.