PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
54/107
Set and Query Motor Analysis Function Channel B Measured Maximum Frequency
Syntax
Command
:EXTernalinB:PULSe:FREQuency <100Hz/500Hz/1kHz/5kHz>
Query
:EXTernalinB:PULSe:FREQuency?
Response
<100Hz/500Hz/1kHz/5kHz>
:Set frequency
Description
Command
Sets the maximum frequency for Channel B during pulse input.
Query
Returns the setting for maximum frequency for Channel B pulse input as a
character string.
Example
Command
:EXTB:PULS:FREQ 500Hz
Set the maximum frequency for Channel B during pulse input to 500Hz.
Query
:EXTB:PULS:FREQ?
Response
:EXTERNALINB:PULSE:FREQUENCY 500Hz (WHEN HEADER IS ON)
500Hz (WHEN HEADER IS OFF)
Note
The setting for this command is common with that of ":AOUT:FREQuency" Command.
Set and Query Motor Analysis Function Motor Pole Value
Syntax
Command
:EXTernalinB:PULSe:MOTorpoles <Pole value>
Query
:EXTernalinB:PULSe:MOTorpoles?
Response
Pole value data :even numbers in 02 - 98
Description
Command
Sets the pole value.
Query
Returns the setting for the pole value.
Example
Command
:EXTB:PULS:MOT 8
Set the motor pole value to 8.
Query
:EXTB:PULS:MOT?
Response
:EXTERNALINB:PULSE:MOTORPOLES 8 (WHEN HEADER IS ON)
08 (WHEN HEADER IS OFF)
Note
When an odd number is selected, an even number smaller than the selected number will be
set.