PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
55/107
Set and Query Pulse Values of Motor Analysis Function Channel B
Syntax
Command
:EXTernalinB:PULSe:NUMBer <Pulse>
Query
:EXTernalinB:PULSe:NUMBer?
Response
Pulse
:00001 to 60000
Description
Command
Sets the pulse for Channel B.
Query
Returns the pulse setting for Channel B as a numerical value.
Example
Command
:EXTB:PULS:NUMB 360
Set the pulse to 360.
Query
:EXTB:PULS:NUMB?
Response
:EXTERNALINB:PULSE:NUMBER 00360 (WHEN HEADER IS ON)
00360 (WHEN HEADER IS OFF)
Note
Only 1/2 multiples of the motor pole values can be set.
Set and Query Motor Analysis Function Channel B Scaling
Syntax
Command
:EXTernalinB:SCALe <Scaling Value>
Query
:EXTernalinB:SCALe?
Response
Scaling Value
:0000.01 to 9999.99
Description
Command
Sets the scaling value for Channel B.
Query
Returns the scaling setting for Channel B as numerical values.
Example
Command
:EXTB:SCAL 2.0
Set the scaling value for Channel B to 2.0.
Query
:EXTB:SCAL?
Response
:EXTERNALINB:SCALE 0002.00 (WHEN HEADER IS ON)
0002.00 (WHEN HEADER IS OFF)
Note
This command is only effective under the :EXTernalinB:PULSe OFF.