PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
51/107
Set and Query Motor Analysis Function
’s Motor Synchronized Sources
Syntax
Command
:EXTernalinA:SOURce < Synchronized Sources >
Query
:EXTernalinA:SOURce?
Response
< Synchronized Sources >
:U1/U2/U3/U4/I1/I2/I3/I4/Ext/DC50ms/DC100ms
Description
Command
Sets the motor synchronized sources.
Query
Returns the motor synchronized Sources as character strings.
Example
Command
:EXTA:SOUR U1
Set the motor synchronized Sources to U1.
Query
:EXTA:SOUR?
Response
:EXTERNALINA:SOURCE U1 (WHEN HEADER IS ON)
U1 (WHEN HEADER IS OFF)
Note
Ext can only be set with key and when CHB is set as the pulse.
Select and Query Motor Analysis Function Channel A Unit
Syntax
Command
:EXTernalinA:UNIT <Unit data>
Query
:EXTernalinA:UNIT?
Response
<Unit data>
:V/Hz/mNm/Nm/kNm
Description
Command
Sets the unit for Channel A.
Query
Returns the unit setting for Channel A as a character string.
Example
Command
:EXTA:UNIT V
Set the Channel A unit to V.
Query
:EXTA:UNIT?
Response
:EXTERNALINA:UNIT V (WHEN HEADER IS ON)
V (WHEN HEADER IS OFF)
Note
During Analog DC setting, when Channel A sets to “Hz”, Channel A input setting changes to
frequency, and during frequency setting, when Channel A sets to “V”, Channel A input setting
changes to Analog DC.