![Hioki MEMORY HiCORDER 8855 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/hioki/memory-hicorder-8855/memory-hicorder-8855_instruction-manual_2137316047.webp)
27
────────────────────────────────────────────────────
2.11 Notes on Measurement
────────────────────────────────────────────────────
DANGER
Maximum input voltage ratings for the input module and the
input terminals of the instrument are shown below. To avoid
the risk of electric shock and damage to the instrument, take
care not to exceed these ratings.
The maximum rated voltage to earth of the input module
(voltage between input terminals and main instrument frame
ground, and between inputs of other analog input modules) is
shown below. To avoid the risk of electric shock and damage
to the instrument, take care that voltage between channels
and between a channel and ground does not exceed these
ratings.
The maximum rated voltage to earth rating applies also if an
input attenuator or similar is used. Ensure that voltage does
not exceed these ratings.
When measuring power line voltages with the 8950, 8952 or
8953-10, always connect the probe to the secondary side of
the circuit breaker. Connection to the primary side involves
the risk of electric shock and damage to the instrument.
Before using the instrument, make sure that the insulation on
the connection cords is undamaged and that no bare
conductors are improperly exposed. Using the products in
such conditions could cause an electric shock, so contact
your dealer or Hioki representative for replacements. (Model
9197 or 9198 CONNECTION CORD.)
Input/output terminal
Maximum input voltage
Maximum rated voltage to earth
8950 (input)
400 V DC max.
370 V AC/DC
8951 (input)
30 V rms or 60 V DC
30 V rms or 60 V DC
8952 (input)
400 V DC max.
370 V AC/
D
C
8953-10 (input)
400 V DC max.
370 V AC/
D
C
8954 (input)
30 V rms or 60 V DC
370 V AC/
D
C
8955 (input)
30 V rms or 60 V DC
30 V rms or 60 V DC
9322
2000 V DC, 1000 V AC (CAT
II)
600 V DC/AC (CAT III)
When using grabber clips
1500 V DC/AC (CAT II), 600 V DC/AC (CAT III)
When using alligator clips
1000 V DC/AC (CAT II), 600 V DC/AC (CAT III)
EXT TRIG/ START
STOP/ EXT SMPL
-5 to +10 V DC
Not insulated
TRIG OUT/ GO/ NG/
EXT.OUT
-20 V to +30 V DC
500 mA max./ 200 mW max.
2.11 Notes on Measurement
Содержание MEMORY HiCORDER 8855
Страница 2: ......
Страница 20: ...xii Chapter Summary...
Страница 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Страница 28: ...8 1 2 Identification of Controls and Indicators...
Страница 113: ...93 4 8 FFT Analysis Function Stored waveform 1 1 octave analysis 1 3 octave analysis Example Octave analysis waveforms...
Страница 116: ...96 4 8 FFT Analysis Function...
Страница 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Страница 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Страница 186: ...166 8 6 Moving Cursors to the Search Points...
Страница 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Страница 263: ......
Страница 264: ......