5.3 Scaling Function (SYSTEM)
────────────────────────────────────────────────────
5.3.2 Scaling Setting Example
0.2
0.1
-0.1
-0.2
5.0
10
-5.0
-10
[
V
]
[
A
]
0
0
Through the use of the scaling function, the
signal from the sensor can be obtained in the
form of a current value.
Cursor values A and B, respectively, show the
current values.
If the gauge is turned on before the printout is
made, the gauge is output in a current value.
The example below shows the type of scaling when the measurement range
is set to 10 A using the 9018 CLAMP ON PROBE.
Scaling method: 2-point method
Setting: SCI or ENG
Converting value: 0.2 (V)
10 (scale), 0 (V)
1
0 (scale)
unit (eu): A