viii
────────────────────────────────────────────────────
Notes on Use
────────────────────────────────────────────────────
DANGER
10:1 and 100:1 probe connections
The maximum rated to-voltage does not change when using a
9665 10:1PROBE or a 9666 100:1PROBE. To avoid electrical
shock or damaging the 8855 instrument, make probe
connections in such a manner that the method for the probe,
and make sure the to-ground voltage does not exceed the
rated maximum.
The maximum input voltage is 1,000 V DC for the 9665
10:1PROBE, and 5,000 V DC for the 9666 100:1PROBE. (The
measurement category (overvoltage category) is the same as
that of the input modules of MEMORY HiCORDERs that use
the 9665 and the 9666. ) Do not measure voltages that exceed
the maximum input voltage, as the 8855 instrument could be
damaged and an accidents resulting in injury or death could
result.
WARNING
Power Supply Connections
Before turning the instrument on, make sure the source
voltage matches that indicated on the instrument's power
connector. Connection to an improper supply voltage may
damage the instrument and present an electrical hazard.
Replacing the Input Modules
To avoid electric shock accident, before removing or
replacing an input module, confirm that the instrument is
turned off and that the connection cords are disconnected.
To avoid the danger of electric shock, never operate the
instrument with an input module removed. To use the
instrument after removing an input module, install a blank
panel over the opening of the removed module.
Grounding the Instrument
To avoid electrical accidents and to maintain the safety
specifications of this instrument, connect the power cord only
to a 3-contact (two-con ground) outlet. Refer to
Section 2.2, "Power Supply and Ground Connections."
Before Powering on
Check that the power supply is correct for the rating of the
instrument. Be careful to avoid connecting voltage improperly,
as the internal circuitry may be destroyed. (The AC fuse is
integrated in the instrument.)
Usage Precautions for the Internal MO Drive (option)
To prevent damage to the instrument, do not attempt to
disassemble the MO drive.
Laser radiation can be emitted when the MO drive is open.
Avoid looking directly into the laser when the MO drive is
open. Maximum laser output is 50 mW (at 685 nm, pulsed).
Содержание MEMORY HiCORDER 8855
Страница 2: ......
Страница 20: ...xii Chapter Summary...
Страница 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Страница 28: ...8 1 2 Identification of Controls and Indicators...
Страница 113: ...93 4 8 FFT Analysis Function Stored waveform 1 1 octave analysis 1 3 octave analysis Example Octave analysis waveforms...
Страница 116: ...96 4 8 FFT Analysis Function...
Страница 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Страница 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Страница 186: ...166 8 6 Moving Cursors to the Search Points...
Страница 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Страница 263: ......
Страница 264: ......