![Hioki MEMORY HiCORDER 8855 Скачать руководство пользователя страница 156](http://html1.mh-extra.com/html/hioki/memory-hicorder-8855/memory-hicorder-8855_instruction-manual_2137316156.webp)
136
────────────────────────────────────────────────────
Chapter 6 Memory Segmentation Function
────────────────────────────────────────────────────
Measurement start
Start block
Measurement end
Trigger mode
SINGLE
REPEAT
AUTO
Waveform display
Auto print
End block
Auto saving
NOTE
Relation between trigger mode and sequential save function
Press the
START
key and LED light.
Data recording starts when trigger condition are met.
Start block is stored by trigger setting.
The following waveform display ON:
waveform is displayed.
The following waveform display OFF:
waveform is not displayed.
End block is stored by trigger setting.
The following waveform display ON:
waveform is displayed.
The following waveform display OFF:
waveform is not displayed.
Displays end block waveform.
When auto printing is enabled, all blocks are printed from the
start block while individual blocks are displayed.
When auto saving is enabled, waveforms in all blocks are saved
in a batch.
REPEAT or AUTO trigger mode returns you to the start block,
and recording starts.
End of measurement
When the
STOP
key is pressed twice during measurement, the
8855 is forcibly stopped. (Auto printout is not executed.)
When auto save is enabled and the 8855 is forcibly stopped
during measurement, the measurement data up to the point
where the 8855 was forcibly stopped is saved automatically.
End of measurement
If the
STOP
key is pressed during recording, recording continues
until the end of the current block.
(Waveform display, auto printout and auto save are executed.)
When the
STOP
key is pressed twice during measurement, the
8855 is forcibly stopped.
When auto save is enabled, waveforms are saved after recording
the end of the current block, and then recording starts from the
start block.
When auto save is enabled and the 8855 is forcibly stopped
during measurement, the measurement data up to the point
where the 8855 was forcibly stopped is saved automatically.
(Auto printout is not executed.)
When the following waveform display (memory function) is on
and the view function is on, the dead time becomes very large.
Содержание MEMORY HiCORDER 8855
Страница 2: ......
Страница 20: ...xii Chapter Summary...
Страница 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Страница 28: ...8 1 2 Identification of Controls and Indicators...
Страница 113: ...93 4 8 FFT Analysis Function Stored waveform 1 1 octave analysis 1 3 octave analysis Example Octave analysis waveforms...
Страница 116: ...96 4 8 FFT Analysis Function...
Страница 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Страница 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Страница 186: ...166 8 6 Moving Cursors to the Search Points...
Страница 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Страница 263: ......
Страница 264: ......