![Hioki MEMORY HiCORDER 8855 Скачать руководство пользователя страница 143](http://html1.mh-extra.com/html/hioki/memory-hicorder-8855/memory-hicorder-8855_instruction-manual_2137316143.webp)
123
────────────────────────────────────────────────────
5.11 Setting the 8954 VOLTAGE/TEMP UNIT
────────────────────────────────────────────────────
Measurement range
10 /DIV
100 /DIV
Measurement input upper limit
200
o
C
2000
o
C
Measurement input lower limit
-100
o
C
-200
o
C
0
1000
-1000
Thermocouple: T
Measurement input range
When the position is 50%, the
measurement range is 100
o
C/DIV, and
the thermocouple measurement input
range (-200 to 400
o
C) is exceeded
400
-200
When the measurement
input range upper limit
is exceeded
When the measurement
input range lower limit is
exceeded
-100
0
-200
When the position is 100%, the measurement
range is 10
o
C/DIV and the waveform is within
the thermocouple measurement input range,
but the 10
o
C/DIV measurement input range
lower limit (-100
o
C) is exceeded
When the measurement
input lower limit
is exceeded
NOTE
Measurement range and measurement input upper and lower limits
Note that the measurement input upper and lower limits change according to
the measurement range.
When the following upper and lower temperature limits are exceeded, the
waveform is saturated.
Differences in the waveform display depending on the
measurement range
(For magnification/compression rate x 1, normal screen, thermocouple: T)
If the input terminals on the 8954 VOLTAGE/TEMP UNIT come into direct
contact with strong winds, the thermal balance of the input module may be
upset, causing an error to occur. When measuring in this type of location,
change the location of the module so that it is not in direct contact with the
wind.
When the peripheral temperature changes suddenly, the thermal balance may be
upset, causing an error to occur. In this case, leave the instrument for
approximately an hour and start measuring once the temperature has stabilized.
Содержание MEMORY HiCORDER 8855
Страница 2: ......
Страница 20: ...xii Chapter Summary...
Страница 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Страница 28: ...8 1 2 Identification of Controls and Indicators...
Страница 113: ...93 4 8 FFT Analysis Function Stored waveform 1 1 octave analysis 1 3 octave analysis Example Octave analysis waveforms...
Страница 116: ...96 4 8 FFT Analysis Function...
Страница 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Страница 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Страница 186: ...166 8 6 Moving Cursors to the Search Points...
Страница 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Страница 263: ......