![Hioki MEMORY HiCORDER 8855 Скачать руководство пользователя страница 145](http://html1.mh-extra.com/html/hioki/memory-hicorder-8855/memory-hicorder-8855_instruction-manual_2137316145.webp)
125
────────────────────────────────────────────────────
5.12 Setting the 8955 F/V UNIT
────────────────────────────────────────────────────
(2) Setting the measurement range
Set the value per division.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Range
item to be
set.
3. Use the
JOG
control, the function keys or the
RANGE
knob to make the selection.
(3) Setting the threshold value
The waveform is measured for the interval that it
crosses the threshold value.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Level
(threshold
value) item to be set.
3. Use the function keys to make the selection.
(4) Setting the pull-up resistance
Pull-up resistance is used when the 8855 is connected to an open-collector
output signal.
When ON is selected, the input terminal is pulled up by +5V. With most
measurements, this option is set to OFF.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Pull Up
item to be set.
3. Use the function keys to make the selection.
(5) Setting the hold function
Displayed measurement results are retained until frequency and rotation
values are verified.
When this is set to OFF and the measured frequency is not verified within
the specified time, the previously displayed value is halved.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Hold
item to be set.
3. Use the function keys to make the selection.
(6) Setting the slope
Set the slope to use as a reference for measurement.
Measurement is made based on the point where the rising or falling slope of
the input signal crosses the threshold.
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Slope
item to be set.
3. Use the function keys to make the selection.
Содержание MEMORY HiCORDER 8855
Страница 2: ......
Страница 20: ...xii Chapter Summary...
Страница 21: ...1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 1 Product Overview...
Страница 28: ...8 1 2 Identification of Controls and Indicators...
Страница 113: ...93 4 8 FFT Analysis Function Stored waveform 1 1 octave analysis 1 3 octave analysis Example Octave analysis waveforms...
Страница 116: ...96 4 8 FFT Analysis Function...
Страница 117: ...97 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 5 Input Channel Settings...
Страница 177: ...157 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Chapter 8 Search Function...
Страница 186: ...166 8 6 Moving Cursors to the Search Points...
Страница 205: ...185 10 5 Example of Printer Output Printing the List MEM Example...
Страница 263: ......
Страница 264: ......