PCL 600
Operators Manual
v 2.0
66
Error number
Error text
Problem/Reference chapter
Printer
014:025:101
Timeout moving printer axis to home.
It has taken too long to home the
printer axis. Check that the switch is
not dirty or damaged. Check that the
motion runs smoothly. Check for a
dip in the air pressure. Try again.
014:025:201
Printer not calibrated.
Home the printer.
014:025:202
Cannot enable printer when printer not calibrated.
014:025:401
Timeout moving printer from material.
It has taken too long to move the
printer from the material. Check that
the switch is not dirty or damaged.
Check that the motion runs
smoothly. Check for a dip in the air
pressure. Try again.
014:025:501
No material found during horizontal printer move.
Make sure there is material between
the printer clamps. Check if the
switches are dirty or damaged.
Check if the motion runs smoothly.
Check for a dip in the air pressure.
014:025:502
No material found during vertical printer move.
014:025:503
Print head not on home position.
Check if the switches are dirty or
damaged. Check if the motion runs
smoothly. Check for a dip in the air
pressure. Try again.
014:025:504
Timeout in sending print head to home position.
014:025:702
Timeout moving under printer to material.
Cutting cell doors
014:027:100
Cell doors unlock took too long, skipping.
The locking mechanism may be
blocked. Check it and try again.
Close the cell doors. If the problem
continues check the door sensors.
014:027:101
Cell doors not closed, continuing to unlock.
014:027:200
Cell doors not closed, unable to lock doors. Close
cell doors.
014:027:300
Cell doors unlocking took too long, continuing to
unlock.
Clamp calibration
014:028:100
Unable to start clamp calibration procedure when
profiling is not stopped.
You cannot calibrate the clamps
during profiling. Press ‘Stop
profiling’ before calibration.
Attention! This will reset the
profiling!
014:028:101
Unable to start clamp calibration procedure when
clamps are not homed.
Home the clamps before starting
this procedure.
014:028:102
Unable to continue clamp calibration procedure
when profiling is not stopped.
Profiling has been started after
beginning the calibration procedure.
. Press ‘Stop profiling’ before
calibration. Attention! This will reset
the profiling!
014:028:104
Unable to continue clamp calibration procedure
when not all clamps are opened.
Open the clamps. Check if the
clamps are running smoothly and
there are no blockages.