PMX command set (API)
PMX
A4354-2.2
HBM: public
443
Explanations:
Glitch Filter
,
p1
Input signals with pulse widths < x μs are not evaluated.
Default Value is 1 (1μs)
.
F2 evaluation
,
p5
Activates detection and evaluation of direction of rotation
using the second Hardware input. Default Value is
0
(off)
.
Resolution quadrupling
,
p6
All edges adjacent to F1 and F2 are evaluated
.
If F2 is
not connected or F2
evaluation is turned off (p6=
0
)
,
there
is a resolution doubling. Default Value is
0
(off)
.
Zero index input
,
p7
Hardware input for the zero index signal
.
Relevant for
incremental transducers. In count mode an active (elec
trical high level) zero index signal sets the counter value
to p1
0.
Default Value is
0.
Factor
,
p8
Factor for automatic resetting the counter value.to p1
0
after p8*p9 pulses (p8 rotations)
.
Default Value is
0.
Pulses per revolution
,
p9
Used for rotary encoders. Pulse count for a single turn
.
If
resolution quadrupling is activated this value must be
multiplied by 2 or 4. Default Value is
0.
The “physical
zero
index” provides the zero position
.
The
“pulses per revolution” setting knows the maximum value
for the transition
zero
³
maximum
.
The “pulses per revolution” in combina
tion with the “physical
zero
index” may also align (com
pensate) a quarter
pulse width difference between these
two competing ways of resetting the counter value
.
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...