Internal calculation channels
PMX
A4354-2.2
HBM: public
213
13.2.4.3 RTD PT100 at PX455
Function
Temperature measurement with a RTD Pt100 resistor
connected at PX455.
Range -100°C ... +500°C.
A fixed 100 Ohms precision resistor must complement the
RTD to form a half bridge. The complement resistor is
connected at the positive bridge supply.
The complement resistor can be connected directly at the
PX455 front clamps without wiring.
The details are outlined in a PMX Tech Note.
Inputs
The half bridge signal in mV/V
Outputs
Temperature in °C
Parameters
R_wire [Ohms]: The resistance of the wire to the RTD
(single distance). This parameter must be set, when the
complement resistor is connected without wires directly at
the PX455 clamps.
If the complement resistor is connected with the same wire
length as the RTD 100, then the wire resistances are
compensated by the bridge layout and R_wire must be
zero.
Default
R_wire = 0 Ohms
Exception handling
If the input is marked as invalid (Invalid bit), the operation
stops and the output gets frozen and is marked as invalid.
When the temperature is out of range (-100 .. +500°C),
then -333.3°C is output and the output is marked as
invalid.
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...