Internal calculation channels
PMX
A4354-2.2
HBM: public
215
13.2.4.4 Signal generators (rectangle, triangle, sine)
Function
Generates a periodic or non‐periodic signal
Inputs
Enable bit (digital input)
Starts the generation. When low, the generator stops and
outputs the offset value. When enable is high again, the
generator does not continue where it stopped but starts with a
new period.
Outputs
Generates signal
out
0
Parameters
-
Wave form
[sine
,
rectangle
,
noise, counter, constants,
ramp]
-
Frequency
0
..
Data rate/4 (default 19200/4 Hz = 4800 Hz)
(only effective with sine
,
rectangle
,
triangle)
-
Amplitude
-
Offset
-
Periods
Number of periods to be played. Only effective with
periodical waveforms. Zero means endless operation.
Maximum value 2
31
-1.
When the periods have been played, the offset value is
output. For restart, the enable input must toggle to low, then
high.
The generation stops immediately when the enable input
gets low, even when the periods were not fully played.
Default
Sine
,
amplitude 1.0, offset 0.0, Frequency 100 Hz, 0 periods
(=endless)
Enable input = constant high
Exception handling
If the range is overrun, NotaNumber is output.
The
noise signal
is generated from a pseudorandom
series of numbers with the period 2
31
.
The
counter
counts upwards with the data rate (default
19,200 Hz)
.
Range: -Amplitude to +Amplitude (with offset = 0).
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...