Internal calculation channels
194
A4354-2.2
HBM: public
PMX
13.2.2.5 Hold function (analog)
Function
Holds an input value controlled by a measured value (or a
calculated channel)
Hold condition: When the control input is inside (or optionally
outside) the
[threshLow
,
threshHigh]
interval.
When the hold condition is met
,
the input in0 is output to the
output out
0
(including a status
bit)
.
When the hold condition is not met
,
the output is held at the
last hold value and may be made invalid
.
Inputs
- Input
in0
- Control input
control
- Digital Init
input (controls initialization
,
high
active)
Outputs
Hold value
out0
Parameters
-
Upper threshold
of hold range
threshHigh
-
Lower threshold
of hold range
threshLow
-
Hold outside
holdOutside
(yes/no)
,
hold outside the interval
[threshLow
,
threshHigh]
-
Only on entry in the interval
onEntryOnly
(yes/no)
yes:
The value is hold on entry in the interval (if applicable,
after expiry of the delay) and then held up to the next entry.
no
: The output is constantly updated (if applicable, after the
delay has expired), as long as the hold condition has been
met
.
-
Delay
in ms
.
Range 0.0
…
60000.0 ms (= 1 min); Resolution 0.1 ms
The value is only held when the hold condition remained
active without interruption for this period
-
Invalid outside
invalidOutside
(yes/no)
The output value is marked as invalid as long as the hold
condition is not met
-
Initial value
initValue
Is output when the edge is positive at the digital Init
input
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...