PMX command set (API)
PMX
A4354-2.2
HBM: public
429
signal source) and electrical (p3, output in Volt) values of
the voltage output. Virtual subchannels (channel 9) do
not support calibration points. It is possible to define
parameters p1, p2 and p3, but this is ignored!
Note:
The response depends on the SRB command.
Calibration Point Query
Output the characteristic points of the transducer (input)
Output the input characteristic points (transducer) of all
the selected channels (PCS/SPS).
Syntax:
CAP?<p1>(x)
Parameters:
p1:Number of points (1 or 2)
Response:
q1,q2,q3(y)
q1: Number of points (1 or 2)
q2: Measurement signal (amplifier-dependent unit)
q3: Value in the displayed units
Virtual subchannels (channel 9) do not support calibration
points.
Response for p1=1: q2=0, q3=0.
Response for p1=2: q2=100, q3=100.
Calibration
Calibrate the amplifier
Calibrate the amplifier, all selected channels (PCS/SPS).
Activates the Cal option, by implication. The ACL setting
does NOT change! Only supported for PX455! Other
(meas‐)cards return “OK”(0).
CAP?
CAL
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...