Diagnosis and maintenance
526
A4354-2.2
HBM: public
PMX
24.7 Replacing measurement and
communication cards
Measurement and communication cards can be
retrofitted or removed at a later date. Please note the
combination options (see Chapter 8.2.1).
After modification and switching on the supply voltage,
the PMX automatically detects and initializes the
hardware configuration.
Notice
If measurement cards are removed after being updated
or mounted in different slots, the factory settings are
loaded. All parameters must be reentered, also for the
existing cards.
If communication cards (EtherCAT, PROFINET or
EtherNet/IP) are updated, replaced or removed, the
parameter settings are retained. Adaptation to the new
fieldbus is necessary here via the configuration tool of the
controller.
Important
A CODESYS application or web visualization running on
the PMX will also be retained after a firmware update or if
a card is changed (firmware 2.00 or higher). Please note
that signal mapping in CODESYS is fixed and if the
measuring cards are moved or the PMX computing chan
nels are changed, the signal mapping must be checked
and corrected if necessary.
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...