PEL-3000(H) User Manual
146
Memo
A user-created note for the
currently selected OCP
function.
Range
High(CC Mode High),
Mid(CC Mode Middle) and
Low(CC Mode Low)
Start Current(Start C) Starting current value for
the test.
End Current(End C) The current value that will
end the test. The value must
be higher than the OCP
value of the DUT you are
testing. This parameter is
used as a fail-safe for if the
over current protection of
the DUT fails. If the
measured current is reaches
End Current value it would
then indicate that the power
supply OCP failed.
Step Current(Step_C) Sets the step resolution of
the current.
Last Current(Last_C) Sets the final current value
after OCP has been tripped.
This is the steady-state
current draw after the OCP
has been tripped.
Step Time(Step_T)
Sets the execution time of
each step. (50ms to 1600s)
Delay Time(Delay)
The OCP testing delay time.
Sets the how long to delay
starting the test after the
Load On key has been
pressed. (5ms ~ 160ms)
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...