PEL-3000(H) User Manual
148
Last_C
Step_T
Delay
Trig_V
4.
Press the
Save[F3]
to save the selected test
setup.
Start OCP
5.
Press
OCP[F1]
to turn the OCP function on if it
is off.
6.
The OCP function can be started by turning the
load on by pressing
Shift
Load
On/
Off
.
The test current will increase from the Start C
value to the End C value in steps according to
the Step C value, until the test has finished.
The test will start running when the power
supply voltage is greater than the Trig V
voltage.
Example: OCP
Function running
OCP
RS232
01/Oct/2017
OCP Test
7.498
V
0.1531
A
1.15
W
Set Curr:
Fetch Volt:
2.000
1.500
1.000
6.91
6.92
6.91
Measured voltage for
last three steps
Measured voltage,
current and power
Running.
Set current for the
last three steps
(descending order)
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...