PEL-3000(H) User Manual
174
Note
Use resistors with minimum residual resistance of
50
Ω
or less.
Note for proportional control: Do not use swtiches
that switch between fixed resistances. Please use
continuously variable resistors.
External Resistance Control – Operation
Description
External resistance control can be used to control
the current, voltage, resistance and power for CC,
CR, CV and CP modes. Configuration for each
operating mode is the same.
CC Mode
Proportional Control:
Input current = rated current × (external
resistance/10).
Inverse Control:
Input current = rated current × (1 - external
resistance/10).
Input
Current
External
Resistance
Rated
Current
0
Ω
10k
Ω
Proportional control
Inverse control
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...