PEL-3000(H) User Manual
56
such as the slew rate, and switching function
settings. See page 65 for details.
Display
Mode
CR
I Range
H 8.75A
V Range
L 80V
Function
Static
Configure
0.000
V
0.0000
A
0.00
W
SlewRate
14.000
CR B Value
0.54422
CR A Value
0.54422
LOAD
RS232
A Value
Fine
01/Oct/2017
mA/uS
Ω
Ω
Mode
Current range
Voltage
range
Active
setting
Conductance/
Resistance
settings
Note
Basic CR mode configuration is complete. See page 65
for more configuration options.
The current range and voltage range applies to all the
operating modes.
CR Units
Description
The CR setting units can be set to ohm (Ω) or
millisiemens (mS)
.
Operation
1.
Make sure the load is off.
2.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
CR Unit
setting.
Range:
Ω
, mS
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...