EXTERNAL CONTROL
181
Pin Inputs
Pins 8 and 9 of the J1 connector are internally
pulled up to 5V with a 10kΩ resistor when open.
When closed, pin 8 and 9 are pulled down to the A
COM ground level.
PEL-3000(H)
Switches
Analog
connector
12
+5V
A COM
8
9
10k
Ω
Note
The range can only be externally controlled when the
IRange has been set to High using the front panel
controls.
I Range Status
Description
Pins 14 and 15 (Range Status 1& 0) of the J1
connector are used to monitor the IRange status.
The pinout combination determines the range
status.
I Range
Pin 14 Pin 15
H
Off
Off
M
Off
On
L
On
Off
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...