PEL-3000(H) User Manual
70
CV, +CV Mode Response Speed
Description
The response speed setting is the response speed
for the negative feedback control of the load
current when used in CV, +CV mode. Response
speed settings are only applicable to CV, +CV
mode.
A response speed that is too fast could cause
the unit to be unstable.
Reducing the response speed can improve
stability.
Operation
1.
Make sure the load is off.
2.
Press
Main
. Make sure the unit is in CV
mode by using the
Mode[F1]
soft-key.
3.
Select the response speed with the
Response[F4]
soft-key.
Range:
Slow, Fast (Fast, 6, 5, 4, 3, 2, 1, slow)
CV mode: The response speed settings
Fast, 6, 5, 4 are all the same
for CV mode.
+CV mode: The response speed settings
5 and 4 are the same for CV
mode. The response speed
settings Slow and 1 is the
same.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...