CC + CV Mode
Load Input
Voltage
Load Current
CV level
CC Current
Value
Note that when the source voltage is less than the
CV level, no current will flow due to a very high
impedance.
CR Mode
CR Mode
When the unit is set to CR mode it will operate as a
constant resistance load when connected to a
constant voltage (CV) or constant current (CC)
source. This means the unit will maintain a set
resistance, up to the rated power, regardless of the
load voltage or current. When input voltage
changes, the unit responds by changing the current
load to maintain the set resistance according to
ohm’s law.
CV source : Load current = Load voltage / CR
setting value
CC source : Load voltage = Load current x CR
setting value
This is illustrated below.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...