FUNCTION MENU
159
IHVL(I range high, V range low)
ILVH(I range low, V range high)
IMVH(I range middle, V range high)
IHVH(I range high, V range high)
Setting
Sets the values corresponding to the
defined discharging mode (CC
mode in A, CR mode in mS and CP
mode in W).
SlewRate
=
Sets the test rising slew rate in
mA/us (not adjustable for CP
mode).
SlewRate
?
Sets the test falling slew rate in
mA/us (not adjustable for CP
mode).
Stop Volt
Sets the voltage at which the test
should be interrupted. The value
must be lower than the battery start
voltage.
Stop Time
Sets the time after which the test
should be interrupted (max value is
999h:59m:59s).
Stop AH
Sets the discharged energy rate at
which the test should be interrupted
(Max value is 9999.99Ah).
Datalog timer Sets the time interval for data
capture. Up to 65,535 data can be
saved when running data logging
function. When logging data reaches
to the maximum amount, it won’t be
saved and be ignored.
Panel operation
1.
Press
FUNC
>
Next Manu[F5].
>
BATT[F2].
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...