EXTERNAL CONTROL
171
See page 55 for CR mode.
See page 57 for CV mode.
See page 58 for CP mode.
7.
Press
Main
>
Configure [F5]
>
Next Menu [F4]
>
External [F3]
.
8.
Set the
Control
parameter to V.
The J1 connector is now ready for external
voltage control.
Adjust offset and full scale with variable resistor (PEL-3021H/
PEL-3041H/ PEL-3111H)
Variable Resistor
in rear panel
FS
OS
FS
OS
VR1
VR2
VR3
VR4
CC/CR/CV/CP
+CV
Operation
CC, CR, CV, CP
Mode
1.
Apply a voltage of 1V to pin J1-1 based on the
level of pin J1-3.
2.
Turn VR2 with screwdriver to adjust the value
to 10% of the rating in each the operating mode.
3.
Apply a voltage of 10V to pin J1-1 based on the
level of pin J1-3.
4.
Turn VR1 with screwdriver to adjust the value
to 100% of the rating in each the operating
mode.
5.
Apply a voltage of 1V to pin J1-1 based on the
level of pin J1-3.
6.
Turn VR2 with screwdriver to adjust the value
to 10% of the rating in each the operating mode.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...