Note
Readjustment is needed when you use a different
operating mode, current range or voltage range.
Cx+CV mode
1.
Apply a voltage of 1V to pin J1-1 based on the
level of pin J1-3.
2.
Turn VR4 with screwdriver to adjust the value
to 10% of the rating in each +CV mode
.
3.
Apply a voltage of 10V to pin J1-2 based on the
level of pin J1-3.
4.
Turn VR3 with screwdriver to adjust the value
to 100% of the rating in each +CV mode.
5.
Apply a voltage of 1V to pin J1-2 based on the
level of pin J1-3.
6.
Turn VR4 with screwdriver to adjust the value
to 10% of the rating in each +CV mode.
Note
Readjustment is needed when you use a different
voltage range.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...