OPERATION
59
6.
Set the power level parameters using the scroll
wheel and number pad.
For Static mode, set
CP A Value
and/or
CP B
Value
.
For Dynamic mode, set
Level1
and
Level2
.
The maximum and minimum power levels
depend on the selected current range.
For static mode, the parameter that is set last
becomes the “active” setting. This will be
shown in the Operation Status Panel.
7.
To add CV mode to CP mode (CP+CV), see
page 60.
8.
Set the remaining basic configuration settings
such as the slew rate, and timer settings. See
page 65 for details.
Display
Mode
CP
I Range
H 8.75A
V Range
H 800V
Function
Static
Configure
0.00
V
0.0000
A
0.00
W
CP B Value
0.00
CP A Value
0.00
LOAD
RS232
A Value
Fine
01/Oct/2017
W
W
Current range
Active
setting
Current setting
Power
settings
+CV
OFF
Mode
Voltage
range
Note
Basic CP mode configuration is complete. See page 65
for more configuration options.
The current range and voltage range applies to all the
operating modes.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...